Delta
Used to measure the difference between the reference point and a
certain point on the trace, including the stimulus delta (X-axis value)
and response delta (Y-axis value). By default, Delta is disabled. The
delta marker value is displayed at the upper-right corner of the window
or displayed in the marker table. The delta marker is displayed as
"Δ+marker number".
- If a reference marker already exists for the current trace, when
"Delta" is enabled, it calculates the difference between the current
active marker and the reference marker. If the reference marker is the
current active marker, then the Delta function is disabled.
- If no reference marker exists on the current trace, when "Delta" is
enabled, create a reference marker in the center position of the trace,
then adjust the positions of the reference marker and the active
marker.

- When you change the position of the Delta marker, the position of
the reference marker remains unchanged, but the frequency (or time)
difference between the two markers will change along with it.
- The frequency (or Power) difference between the two markers and the
measurement values difference Δ are displayed on the screen.
Discrete
Used to display values at only the discrete points where data is
measured.
- OFF: By default, it is disabled. The marker readouts of all the data
points measured on the trace are displayed.
- ON: Displays values at only the discrete points where data is
measured. The marker readouts of data values that are interpolated from
the data points will not be displayed. The discrete points where data is
measured are determined by Sweep
Points.
Type
- Normal: The marker has a fixed stimulus position
(X-axis value), and by default, the response position (Y-axis value) of
the marker changes with the trace data.
- Fixed: Causes the marker to have a fixed X-axis and
Y-axis position based on its placement on the trace when it was set to
fixed. It does NOT move with trace data amplitude. It can be scrolled
left and right on the X-axis by changing the marker stimulus value.
Format
The newly created marker use the trace data format by default. If you
want to modify the data format of the marker, the following formats are
available.
Trace Default, Linear Magnitude, Log Magnitude, Phase, Group Delay,
Real, Imaginary, SWR, Log/Phase, Lin/Phase, Real/Imag, R + jX, and G +
jB. For details, refer to Data Format.
Coupling Mode
The coupled marker function cause markers on different traces to line
up with the markers on the selected trace. When you move the specified
marker, all the other markers that are coupled with the marker will be
moved together.
Markers are coupled by marker number. If the x-axis domain of
different traces is the same (e.g., frequency or power), then coupling
occurs to the markers with the same marker number on these traces. Trace
markers in a different x-axis domain will not be coupled.
The available coupling modes include:
- OFF: disables the marker coupling.
- Channel: By default, a marker on one trace is
coupled to the same numbered markers on traces which share the same
channel number as the original trace.
- All: A marker on one trace is coupled to the same
numbered markers on all channels, all windows, and all traces.
Marker Table
- OFF: By default, it is disabled. The marker table will not be
displayed.
- ON: when it is enabled, the marker table displaying all the marker
data on the current active trace is displayed below the trace window,
including the marker number, stimulus value, and response value.

All Off
Click or tap All Off to disable the display of all
the markers on the current active trace. The coupled markers are also
disabled.