Main

The following table lists the S-parameter calibration type supported by VNA.

Number of Ports Calibration Type Calibration Item to be Performed
1

(Only calibrate the reflection parameters)

Response Open Open response calibration
Response Short Short response calibration
OSL Open, Short, and Load response calibration
2 Response 1 to 2 Port 1→2 Through response calibration
Response 2 to 1 Port 2→1 Through response calibration
Enh Response 1 to 2 Enhanced response calibration
  • Port 1: Open, Short, and Load
  • Port 1→2: Through
Enh Response 2 to 1 Enhanced response calibration
  • Port 2: Open, Short, and Load
  • Port 2→1: Through
SOLT All response calibration
  • Port 1: Open, Short, and Load
  • Port 2: Open, Short, and Load
  • Port 1⇄2: Through