All the settings are configured for the channel, and take effect for all the traces of the current channel.
Measure Port
Specifies the test port to perform the auto port extension and controls the on/off status of the port, which is updated synchronously as what "Select Ports" displays.
By default, APE feature is enabled for all the test ports automatically. The Auto Port Extensions settings apply to all measurement items associated with the port on the active channel.
Include Losses
By default, it is disabled and the loss is automatically set to 0; the "Adjustment for Mismatch" is also grayed out and disabled.
When it is enabled, the system automatically calculates the loss parameters and writes the port extension configuration to compensate for the loss of transmission line/fixtures (including DC loss and frequency-related loss).
Adjustment for Mismatch
By default, it is disabled. It is only available to set when "Include Losses" is enabled. When you check the checkbox of "Adjustment for Mismatch", the measurement trace will be adjusted, forcing the amplitudes of all data points ≤ 0 dB, avoiding instability of S-parameter values in the simulation.
Frequency Span Mode
Selects the frequency span range used in the calculation of APE parameters. Three modes are available to choose:
User Span
It is only available to set when you select "User Span" under the "Frequency Span Mode". It is used to set a user-defined frequency span.