Port Extension

Port extensions allow you to electrically move the measurement reference plane after you have performed a calibration. After completing the S-parameter calibration, use the port extension to address the following problems in the test scenarios.

Click or tap Calibration > Port Extension > Port Extensions to enter the Port Extensions interface.

Configure the port extensions parameters

Procedures for performing port extension

  1. Select a calibrated S11 measurement trace. Take reflection coefficient of Port 1 as the basic measurement.
  2. Set the data format of the measurement trace to "Phase" to better observe and adjust the phase.
  3. Connect Open or Short at the calibration reference plane. View and verify the phase readouts across the frequency range. In the ideal state, the phase shall be at or near zero. If the phase deviation is large, check whether there is a problem in the calibration process or hardware connection.
  4. Connect the fixture or added transmission line to the measurement system in place of the DUT. In the position where DUT is located, attach an Open or Short.
  5. Adjust the delay compensation
    • If the fixture or the "electrical length" of additional transmission line is known, set the "Delay" parameter directly.
    • If the fixture or the "physical length" of additional transmission line is known, set the "Distance" parameter directly.
    • If the extension length is unknown, increase the value of "Delay" or "Distance" gradually to observe the phase response within the frequency span shown on the screen. Adjust the value continuously until the phase response within the specified frequency span becomes smooth, that is, ensure that the phase variation is within the acceptable range. Then complete the port extension operation.

After performing the above operations, observe whether the phase, reflection coefficient, and other parameters at different frequency points meet the expected values. Compare the results with the results that have not performed port extension to confirm whether the setting is valid. If the measurement results are not ideal, repeat the above steps and readjust the relevant parameters.