Group Delay
Group delay is the time delay of the amplitude envelope of a signal as it passes through a device or network. It represents the rate of change of phase shift with respect to frequency, indicating how different frequency components within a signal experience different transmission delays through the device. The group delay is a measure of phase distortion. It describes the transit time of a signal through a device versus frequency. If the group delay of a system is a constant value for all the frequencies, it indicates that the transmission speed of the frequency components is the same, which will not cause signal distortion.
The phase characteristics of a device typically consists of both linear and higher order (deviations from linear) phase-shift components.
For a linear time-invariant (LTI) system, the relationship between input and output can be characterized by its frequency response.
τ(ω) = -dϕ(ω)/ dω, wherein, ω indicates the angular frequency, Φ(ω) Indicates the frequency response;
τ(f) = -dθ/ df, wherein, θ indicates phase angle; f indicates the frequency.
Tip:
The group delay function is only valid when the data format is Group Delay. The setting is only valid for the current active trace.
Group Delay Aperture
Group delay aperture typically refers to the frequency resolution of the measurement system in group delay measurements.
During a group delay measurement, phase is measured at two closely spaced frequencies and then computes the phase slope. The frequency interval (frequency delta) between the two phase measurement points is called the aperture. Changing the aperture can result in different values of group delay. The computed slope (-delta phase / delta frequency) varies as the aperture is increased. Therefore, when you compare group delay data, you must know the aperture that was used in measurements.
If the group delay aperture is set too large, it may lead to over-smoothing of the measurement results, thereby missing some important phase variation details; if set too small, it may result in increased measurement noise, affecting the stability and accuracy of the measurement results.
The number of adjacent data points can be set using any of the following methods. The three options set the aperture from different angles. Modifying one parameter will affect the other two parameters.
Formula: (Aperture Points - 1) / (Sweep Points - 1) = Percent of Span
(Sweep Stop Frequency - Sweep Start Frequency) / (Sweep Points - 1) = Sweep Points Interval
Sweep Points Interval * (Aperture Points - 1) = Aperture Frequency
When the frequency span or number of points is reduced so that the current Group Delay Aperture is not attainable, the Aperture is adjusted to the new frequency span or number of points.