Calibration interpolation adjusts calibration error terms to match changes to the stimulus parameters settings (such as freq and points) that you make after a calibration is performed or a CalSet is applied, keeping the accuracy of the measurement results through the interpolation algorithm, avoiding error calibration failure arising from modifying the settings.
Enable or Disable the Interpolation
By default, interpolation is disabled. When enabled, the interpolation is enabled for the current active measurement, but this does not necessarily mean that the measurement is interpolated. When enabled, if you modify the following stimulus settings, and the system deems that it is necessary to initiate interpolation operation to keep the accuracy of the measurement, then the interpolation will be applied. If you modify the stimulus settings while the interpolation is disabled, then interpolation is not applied, but the error correction function (Correction/Correction Properties) is automatically disabled.
Interpolation occurs (if enabled) if you change the following settings in Sweep.
Tip:
Decreasing the start frequency or increasing the stop frequency will force the correction function to be disabled.
Interpolation Calibration Accuracy
When a measurement is interpolated, the accuracy of the measurements cannot be predicted. It may be affected significantly or not at all. Identifying measurement errors in these cases must be determined on a case-by-case basis.
When the interpolation algorithm encounters an abrupt or large change in the response magnitude or phase, such as can occur at band crossings, large interpolation errors can be included in the displayed data. These errors can be seen as steps or spikes. If this occurs, attempt to use the following methods to avoid it.
Display State of Interpolation Calibration
After executing the interpolation correction, the calibration status displayed in System Status Bar below the window changes. The character C following a star "*" appears. For example, "C* 2-Port".
Interpolation Correction:
If you modify the Sweep points, increase the start frequency or decrease the stop frequency, the interpolation correction is enabled, then the correction state displayed in System Status Bar below the window shows "C*", for example, "C* 2-Port".
Configuration Modification:
If one or multiple stimulus parameters such as IF BW, Sweep Time, or Port Power change, the correction status shows "CΔ". For example, "CΔ 2-Port".