Basic Calibration

In the calibration wizard, select the desired S-parameter calibration type and configure the relevant parameter settings to perform the S-parameter user-defined calibration.

Click or tap Basic Cal to enter the basic calibration interface.

After completing the above configuration, click or tap Next to enter the following interface. If you haven't set the parameters well, you can click or tap Back to go to the previous step to configure the parameters.

In the menu, you can view the currently selected port to be calibrated and the calibration items to be performed on the specified port. The calibration items include OPEN-M- (Open calibration measurement), SHORT-M- (Short calibration measurement), BROADBAND LOAD-M- (Load calibration measurement), and THRU (Through calibration measurement). Follow the procedures below to perform the basic calibration.

  1. Connect the calibration kits to the test ports of VNA properly according to the pop-up messages.
  2. Click or tap the first calibration button, then the current calibration trace window is displayed below the window. When "√" (a tick icon) is displayed on the button, it indicates that this calibration item is completed.
  3. Repeat Step1 and Step 2 until all the selected calibration items have been calibrated successfully.

  1. To perform the Isolation Calibration, the isolation calibration interface, as shown below.