In the calibration wizard, select the desired S-parameter calibration type and configure the relevant parameter settings to perform the S-parameter user-defined calibration.
Click or tap Basic Cal to enter the basic calibration interface.

Connector
Selects the desired connector type from the drop-down list to select APC 3.5, APC 2.4, APC 7, 7-16, Type F(75), Type N(75), or Type N(50).
Cal Kit
Selects the desired calibration kit model from the drop-down list based on the currently selected connector.
If you check the checkbox of "Show all cal kits", then all of the calibration kits are available to choose without being limited to the connector type.
Port
Check the checkbox of the specified port number to be calibrated, then select the port type (Male or Female) of the DUT that connects the specified port. You can select one or two ports (Port 1 and Port 2) to calibrate.
Cal Type
Selects the desired calibration type from the drop-down list based on the currently selected port. For details, refer to Table 1.
After completing the above configuration, click or tap Next to enter the following interface. If you haven't set the parameters well, you can click or tap Back to go to the previous step to configure the parameters.

In the menu, you can view the currently selected port to be calibrated and the calibration items to be performed on the specified port. The calibration items include OPEN-M- (Open calibration measurement), SHORT-M- (Short calibration measurement), BROADBAND LOAD-M- (Load calibration measurement), and THRU (Through calibration measurement). Follow the procedures below to perform the basic calibration.
Repeat Step1 and Step 2 until all the selected calibration items have been calibrated successfully.
Click or tap Save Cal Data enter the file management interface (File Management). The calibration configuration information of the current channel is saved to the specified path as a file suffixed with "*.cal". Click or tap Finish to complete this calibration, and you will exit from the current interface and go back to the main interface of "Basic Cal". To cancel the calibration, click or tap Cancel.
To perform the Isolation Calibration, the isolation calibration interface, as shown below.
Click or tap all the isolation calibration buttons to perform calibration. After completing the calibration, the calibration configuration file is saved to the specified path, with the filename suffixed with "*.cal". Click or tap Finish to complete the calibration.