In high-speed differential systems, the time difference or length difference between two identical source signals in the transmission path is called signal skewing.
The skew test is based on Time Domain Reflectometry (TDR) technology and accurately quantifies the skew value by transmitting a low-pass step signal to the DUT, comparing the step response time of the reference trace with the measured trace, which can be used for the relative deviation calibration of differential signals.
To use the skew test feature, you need to install the TDA option first.
To configure the skew test parameters, click or tap Skew Measure in the "Math" menu.
Skew State
By default, it is disabled. When it is enabled,
Measure Threshold
It is used to set the measurement point for step voltage. By default, it is 50%. It means measurements are taken at 50% of step voltage by default. Its range is from 10% to 90%. It will automatically adjusted to its boundary value when the value you set is out of its range.
Skew Limit
By default, it is 0 s. Its range is from -1.0e+15 s to 1.0e+15 s.
If the absolute value of the time skew exceeds the limit value, the test fails, displaying "Fail" following the test result; otherwise it shows "Pass".
Reference Trace
The reference trace is used to compare with the current trace, which must be in the same channel as the current trace. To select the reference trace, you need to enable the skew state.