Preparation
Complete the S-parameter calibration for the channel to verify that the system meets the requirements for APE.
Configure Parameters
Set the APE-related parameters in the APE interface, such as target test port, loss, and frequency span.
Connect the Standard Components
Connect Open or Short to the enabled port securely.
Trigger the Measurement
Click or tap "Measure OPEN" or "Measure SHORT". The system verifies the pre-conditions before starting the measurement. If the pre-conditions are met, the measurement will be initiated.
System Calculation and Application
The system automatically checks the history measurement records and current configuration to select the desired calculation method.
After the system calculation is completed, the system automatically activates the port extension function, and the calculated compensation parameters such as delay, loss etc. are applied to the measurement. The trace is updated in a real-time manner. The compensated measurement results will be presented synchronously.
Verify Compensation Effects
Set the data format to Phase. If the phase curve in the full range is flat, it indicates that the port extension compensation is completed. The reference plane has been moved to the position where the standard component is connected.