ECal is a solid-state calibration solution. Every ECal module contains electronic standards that are automatically switched into position during a measurement calibration. These electronic standards have been measured at the factory and the data are stored within the memory of the ECal module. The VNA analyzer uses the stored data, along with the measured data, to calculate the error terms for a measurement calibration.
For the Ecal modules supported by VNA and its operation and connection methods, refer to ECalModule User Guide.
Connect the ECal module port to the port of the analyzer. The Ecal module is controlled and powered through a USB connection. Wear a grounded wrist strap when making ECal calibration. Do not disconnect USB devices and test port during ECal calibration. After connecting the ECal module properly, wait for completing warm-up until the indicator turns green, then start to calibrate.
The procedures for calibration operation are as follows:
Enter the "Select Calibration Ports and ECal Module" interface. Set the following configuration.
Select the calibration type
Supports single port calibration or multiple ports (e.g. 2-Port) calibration. You can select the desired port based on your DUT.
Port Ecal configuration
Selects the desired calibration port based on the calibration type. For 2-port calibration, you must select two different ports.
ECal properties
After connecting properly, VNA can automatically recognize the ECal module and obtain the module information.
Click or tap Next to enter the following interface.
Complete the calibration operation according to the calibration wizard.
After completing the measurement, the following interface is displayed. Click or tap Save As to save the calibration configuration file suffixed with "*.cal" to the specified path. Click or tap Finish to complete the calibration.
